6.5med
CVE-2019-25720
Dräger SC Monitoring Devices DoS via Malformed Network Packet
Published 5 mo ago · modified 3 mo ago · JSON
CVSS
6.5
v3.1 · CNA
CISA KEV
No
not known exploited
EPSS
0.2%
percentile 10%
Public exploits
0
none known
Summary
Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.
CISA SSVC assessment
- Exploitation
- none
- Automatable
- no
- Technical impact
- partial
- Assessed
- 5 mo ago
Exploit prediction (EPSS)
0.20% probability of exploitation in the next 30 days, higher than 10% of all scored CVEs.
CVSS scores
| Assigned by | Version | Score | Vector |
|---|---|---|---|
| CNA | 4.0 | 7.1 | CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N |
| CNA | 3.1 | 6.5 | CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H |
| CNA | 4.0 | 7.1 | CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X |
| CNA | 3.1 | 6.5 | CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H |
Affected products
- Dräger SC 6002XL
- Dräger SC6802XL
- Dräger SC 7000
- Dräger SC8000
- Dräger SC90000 XL
Classification
- Weakness
- CWE-1286 Improper Validation of Syntactic Correctness of Input
References
- static.draeger.com /security/download/2019-11-27-Draeger-SC7000-SC9000-security-advisory-update-v1-5.pdf · vendor-advisory
- vulncheck.com /advisories/dr-ger-sc-monitoring-devices-dos-via-malformed-network-packet · third-party-advisory
Sources
- CVE List: CVE-2019-25720
- NVD: CVE-2019-25720
vulnwire keeps one entry per vulnerability: the CVE id wins, then GHSA; records that alias each other are merged. Machine-readable: JSON.