Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions.

Published 7 y ago · modified 4 mo ago · JSON

CVSS
6.5
v3.0 · NVD
CISA KEV
No
not known exploited
EPSS
0.8%
percentile 56%
Public exploits
0
none known

Summary

Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Log files are accessible over an unauthenticated network connection. By accessing the log files, an attacker is able to gain insights about internals of the patient monitor, the location of the monitor, and wired network configuration.

Exploit prediction (EPSS)

0.78% probability of exploitation in the next 30 days, higher than 56% of all scored CVEs.

CVSS scores

Assigned byVersionScoreVector
NVD3.06.5CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
NVD2.03.3AV:A/AC:L/Au:N/C:P/I:N/A:N

Affected products

  • ICS-CERT Dräger Infinity Delta

Classification

Weakness
  • CWE-532 Insertion of Sensitive Information into Log File