6.2med
CVE-2018-14801
In Philips PageWriter TC10, TC20, TC30, TC50, TC70 Cardiographs, all versions prior to May 2018, an attacker with both the superuser passwo…
Published 8 y ago · modified 4 mo ago · JSON
CVSS
6.2
v3.0 · NVD
CISA KEV
No
not known exploited
EPSS
0.4%
percentile 35%
Public exploits
0
none known
Summary
In Philips PageWriter TC10, TC20, TC30, TC50, TC70 Cardiographs, all versions prior to May 2018, an attacker with both the superuser password and physical access can enter the superuser password that can be used to access and modify all settings on the device, as well as allow the user to reset existing passwords.
Exploit prediction (EPSS)
0.41% probability of exploitation in the next 30 days, higher than 35% of all scored CVEs.
CVSS scores
| Assigned by | Version | Score | Vector |
|---|---|---|---|
| NVD | 3.0 | 6.2 | CVSS:3.0/AV:P/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H |
| NVD | 2.0 | 7.2 | AV:L/AC:L/Au:N/C:C/I:C/A:C |
Affected products
- Philips PageWriter TC10, TC20, TC30, TC50, TC70 Cardiographs
Classification
- Weakness
- CWE-798 Use of Hard-coded Credentials
References
- ics-cert.us-cert.gov /advisories/ICSMA-18-228-01 · x_refsource_MISC, x_transferred
- usa.philips.com /healthcare/about/customer-support/product-security · x_refsource_CONFIRM, x_transferred
- securityfocus.com /bid/105103 · vdb-entry, x_refsource_BID, x_transferred
Sources
- CVE List: CVE-2018-14801
- NVD: CVE-2018-14801
vulnwire keeps one entry per vulnerability: the CVE id wins, then GHSA; records that alias each other are merged. Machine-readable: JSON.