6.2med

CVE-2018-14801

In Philips PageWriter TC10, TC20, TC30, TC50, TC70 Cardiographs, all versions prior to May 2018, an attacker with both the superuser passwo…

Published 8 y ago · modified 4 mo ago · JSON

CVSS
6.2
v3.0 · NVD
CISA KEV
No
not known exploited
EPSS
0.4%
percentile 35%
Public exploits
0
none known

Summary

In Philips PageWriter TC10, TC20, TC30, TC50, TC70 Cardiographs, all versions prior to May 2018, an attacker with both the superuser password and physical access can enter the superuser password that can be used to access and modify all settings on the device, as well as allow the user to reset existing passwords.

Exploit prediction (EPSS)

0.41% probability of exploitation in the next 30 days, higher than 35% of all scored CVEs.

CVSS scores

Assigned byVersionScoreVector
NVD3.06.2CVSS:3.0/AV:P/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H
NVD2.07.2AV:L/AC:L/Au:N/C:C/I:C/A:C

Affected products

  • Philips PageWriter TC10, TC20, TC30, TC50, TC70 Cardiographs

Classification

Weakness
  • CWE-798 Use of Hard-coded Credentials

References

  • ics-cert.us-cert.gov /advisories/ICSMA-18-228-01 · x_refsource_MISC, x_transferred
  • usa.philips.com /healthcare/about/customer-support/product-security · x_refsource_CONFIRM, x_transferred
  • securityfocus.com /bid/105103 · vdb-entry, x_refsource_BID, x_transferred

Sources