CVE-2017-12722
An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6.
Published 8 y ago · modified 4 mo ago · JSON
CVSS
5.3
v3.0 · NVD
CISA KEV
No
not known exploited
EPSS
2.4%
percentile 84%
Public exploits
0
none known
Summary
An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump reads memory out of bounds, causing the communications module to crash. Smiths Medical assesses that the crash of the communications module would not impact the operation of the therapeutic module.
Exploit prediction (EPSS)
2.42% probability of exploitation in the next 30 days, higher than 84% of all scored CVEs.
CVSS scores
| Assigned by | Version | Score | Vector |
|---|---|---|---|
| NVD | 3.0 | 5.3 | CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:L |
| NVD | 2.0 | 5.0 | AV:N/AC:L/Au:N/C:N/I:N/A:P |
Affected products
- n/a Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump
Classification
- Weakness
- CWE-125 Out-of-bounds Read
References
- ics-cert.us-cert.gov /advisories/ICSMA-17-250-02A · x_refsource_MISC, x_transferred
- securityfocus.com /bid/101252 · vdb-entry, x_refsource_BID, x_transferred
- securityfocus.com /bid/100665 · vdb-entry, x_refsource_BID, x_transferred
Sources
- CVE List: CVE-2017-12722
- NVD: CVE-2017-12722