6.5med

CVE-2017-12716

Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communi…

Published 8 y ago · modified 4 mo ago · JSON

CVSS
6.5
v3.0 · NVD
CISA KEV
No
not known exploited
EPSS
0.2%
percentile 15%
Public exploits
0
none known

Summary

Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.

Exploit prediction (EPSS)

0.24% probability of exploitation in the next 30 days, higher than 15% of all scored CVEs.

CVSS scores

Assigned byVersionScoreVector
NVD3.06.5CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
NVD2.03.3AV:A/AC:L/Au:N/C:P/I:N/A:N

Affected products

  • Abbott Laboratories Accent and Anthem

Classification

Weakness
  • CWE-311 Missing Encryption of Sensitive Data
  • CWE-319 Cleartext Transmission of Sensitive Information

References

Sources

vulnwire keeps one entry per vulnerability: the CVE id wins, then GHSA; records that alias each other are merged. Machine-readable: JSON.