CVE-2017-12716
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communi…
Published 8 y ago · modified 4 mo ago · JSON
Summary
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Exploit prediction (EPSS)
0.24% probability of exploitation in the next 30 days, higher than 15% of all scored CVEs.
CVSS scores
| Assigned by | Version | Score | Vector |
|---|---|---|---|
| NVD | 3.0 | 6.5 | CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N |
| NVD | 2.0 | 3.3 | AV:A/AC:L/Au:N/C:P/I:N/A:N |
Affected products
- Abbott Laboratories Accent and Anthem
Classification
References
- ics-cert.us-cert.gov /advisories/ICSMA-17-241-01 · x_refsource_MISC, x_transferred
- securityfocus.com /bid/100523 · vdb-entry, x_refsource_BID, x_transferred
Sources
- CVE List: CVE-2017-12716
- NVD: CVE-2017-12716
vulnwire keeps one entry per vulnerability: the CVE id wins, then GHSA; records that alias each other are merged. Machine-readable: JSON.