9.9crit
CVE-2016-8355
An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0; 3.0; and 3.1.
Published 9 y ago · modified 4 mo ago · JSON
CVSS
9.9
v3.0 · NVD
CISA KEV
No
not known exploited
EPSS
1.5%
percentile 74%
Public exploits
0
none known
Summary
An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0; 3.0; and 3.1. CADD-Solis Medication Safety Software grants an authenticated user elevated privileges on the SQL database, which would allow an authenticated user to modify drug libraries, add and delete users, and change user permissions. According to Smiths-Medical, physical access to the pump is required to install drug library updates.
Exploit prediction (EPSS)
1.49% probability of exploitation in the next 30 days, higher than 74% of all scored CVEs.
CVSS scores
| Assigned by | Version | Score | Vector |
|---|---|---|---|
| NVD | 3.0 | 9.9 | CVSS:3.0/AV:N/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H |
| NVD | 2.0 | 9.0 | AV:N/AC:L/Au:S/C:C/I:C/A:C |
Affected products
- n/a Smiths-Medical CADD-Solis Medication Safety Software through 3.1
Classification
- Weakness
- CWE-306 Missing Authentication for Critical Functionality
References
- ics-cert.us-cert.gov /advisories/ICSMA-16-306-01 · x_refsource_MISC, x_transferred
- securityfocus.com /bid/94630 · vdb-entry, x_refsource_BID, x_transferred
Sources
- CVE List: CVE-2016-8355
- NVD: CVE-2016-8355
vulnwire keeps one entry per vulnerability: the CVE id wins, then GHSA; records that alias each other are merged. Machine-readable: JSON.